The innovation engine for new materials

GPC - chloroform

GPC Chloroform

Gel Permeation Chromatography (GPC) is a separation method for the determination of molecular weight averages (Mn) and molecular weight distribution (PDI=Mw/Mn). The testing report can include calibration plot, chromatogram, number average molecular weight and PDI of the polymer. GPC-chloroform is using chloroform as a solvent. Gel Permeation Chromatography (GPC) separates molecules on the basis of hydrodynamic size (not Mw). The large molecules move more rapidly through the GPC column, and in this way the mixture can be separated. The elution volume of a molecule is related to its hydrodynamic volume. To relate the elution volume to the molecular weight of a polymer, a calibration with molecular weight standards is needed.

For GPC Basics, please read the Waters Beginner's Guide to Size Exclusion Chromatography

Equipment
Pump: Waters Alliance HPLC System, e2695 Separation Module.
Two columns: Tosoh TSKgel SuperHZM-N + guard (MW range: 200-700,000 g/mol)
Detectors:
Waters 2410 Differential Refractometer (RI)
Waters 2998 Photodiode Array Detector (PDA)
Solvent: Chloroform with 0.25%  TEA
Flow Rate: 0.35 mL/min
Injection: 10-20 μL.

Lab: 

Location: 

MRL room 1043

Manufacturer's Info: 

Waters Corporation
43 Maple Street
Milvord, MA 01757
Telephone: 800-252-4752
Internet - http://www.waters.com

Instrument Information and/or Manual: 

AttachmentSize
PDF icon GPC Chloro Manual 2019.pdf153.73 KB

Contact: 

Please contact Rachel Behrens (rachel [at] mrl [dot] ucsb [dot] edu) to schedule training. Before training starts please read the manual and the Waters GPC Primer: https://www.waters.com/nextgen/us/en/education/primers/beginners-guide-t...